Einsatz der Fließ-Injektions-Analyse in markierungsfreien Bindungsassays

DSpace Repositorium (Manakin basiert)

Zur Kurzanzeige

dc.contributor.advisor Gauglitz, Günter de_DE
dc.contributor.author Länge, Kerstin de_DE
dc.date.accessioned 2001-01-04 de_DE
dc.date.accessioned 2014-03-18T10:08:37Z
dc.date.available 2001-01-04 de_DE
dc.date.available 2014-03-18T10:08:37Z
dc.date.issued 2000 de_DE
dc.identifier.other 088971112 de_DE
dc.identifier.uri http://nbn-resolving.de/urn:nbn:de:bsz:21-opus-2040 de_DE
dc.identifier.uri http://hdl.handle.net/10900/48138
dc.description.abstract Im Rahmen dieser Arbeit wurde die Fließ-Injektions-Analyse (FIA) zur automatisierten Handhabung flüssiger Proben mit der Reflektometrischen Interferenz-Spektroskopie (RIfS) als markierungsfreie optische Detektionsmethode kombiniert, um biomolekulare Wechselwirkungen zu untersuchen und im analytischen Bereich anzuwenden. Zunächst wurden Antigen/Antikörper-Systeme mit Hilfe kinetischer und thermodynamischer Konstanten charakterisiert, wobei verschiedene Auswertemethoden der aus der Messung erhaltenen Bindungskurven miteinander verglichen wurden. Des weiteren wurden mit Antigen/Antikörper-Systemen Bindungshemmtests durchgeführt. Dies ist eine Art Immunoassay, mit dem u. a. Antigenkonzentrationen in wäßrigen Lösungen bestimmt werden können. Da der Arbeitsbereich bei hochaffinen Antigen/Antikörper-Systemen unter einer Größenordnung liegt, wurden mit Hilfe des FIA-Systems Bindungshemmtests mit Gradientenansatz eingesetzt, um diesen Bereich zu vergrößern. Damit konnte der Arbeitsbereich sowohl zu höheren als auch zu niedrigeren Antigenkonzentrationen hin erweitert werden: Er betrug nun bis zu drei Größenordnungen beim gleichen Antigen/Antikörper-System. Außerdem wurden Titandioxid-Oberflächen, die als Modellsystem für dentale Titan-Implantate ausgewählt wurden, durch kovalent gebundene Glykoproteine funktionalisiert. Anschließend wurde mit FIA und RIfS die Adsorption von Proteinen auf diesen biofunktionalisierten Schichten untersucht, um diese Oberflächen hinsichtlich einer späteren Verwendung als Implantatbeschichtung zu beurteilen. de_DE
dc.description.abstract Within this doctorate thesis flow injection analysis (FIA) for automated handling of liquid samples was combined with reflectometric interference spectroscopy (RIfS) as a label free optical detection method. This set-up was applied for investigating biomolecular interactions and for using these interactions in the field of analytical measurements. First antigen/antibody-systems were characterised with kinetic and thermodynamic constants, whereas different evaluation methods of the binding curves obtained from the measurements were tested. Afterwards binding inhibition assays were performed with antigen/antibody-systems. These assay type is a kind of immunoassay which enables, among other things, the determination of antigen concentrations in aqueous solutions. Since the working range using high affinity antigen/antibody-systems amounts to less than one order of magnitude, binding inhibition assays with gradient approach were performed with the FIA-system in order to enlarge this range. The working range could be increased both to higher and to lower antigen concentrations: It was now about three orders of magnitude with the same antigen/antibody-system. Furthermore titanium dioxide surfaces, which were chosen as a model system for dental titanium implants, were functionalised with covalent bound glycoproteins. Afterwards the adsorption of proteins on these biofunctionalised surfaces was investigated with FIA and RIfS for rating these surfaces regarding further applications as implant coating. en
dc.language.iso de de_DE
dc.publisher Universität Tübingen de_DE
dc.rights ubt-nopod de_DE
dc.rights.uri http://tobias-lib.uni-tuebingen.de/doku/lic_ubt-nopod.php?la=de de_DE
dc.rights.uri http://tobias-lib.uni-tuebingen.de/doku/lic_ubt-nopod.php?la=en en
dc.subject.classification Fließinjektionsanalyse , Gradientenverfahren , Immunoassay , Implantat de_DE
dc.subject.ddc 540 de_DE
dc.subject.other Bindungsassay , Affinität , Geschwindigkeitskonstanten , Biofunktionalisierung de_DE
dc.subject.other Binding assay , Affinity , Rate constants , Biofunctionalisation en
dc.title Einsatz der Fließ-Injektions-Analyse in markierungsfreien Bindungsassays de_DE
dc.title Application of flow injection analysis in label free binding assays en
dc.type PhDThesis de_DE
dc.date.updated 1970-01-01 de_DE
dcterms.dateAccepted 2000-12-20 de_DE
utue.publikation.fachbereich Sonstige - Chemie und Pharmazie de_DE
utue.publikation.fakultaet 7 Mathematisch-Naturwissenschaftliche Fakultät de_DE
dcterms.DCMIType Text de_DE
utue.publikation.typ doctoralThesis de_DE
utue.opus.id 204 de_DE
thesis.grantor 14 Fakultät für Chemie und Pharmazie de_DE

Dateien:

Das Dokument erscheint in:

Zur Kurzanzeige