Vergleich der Ergebnisse aus Untersuchungen mit dem Tübinger Mobilkampimeter (TMC) und dem Octopus 101 Perimeter an Patienten mit fortgeschrittenen Gesichtsfeldausfällen

DSpace Repositorium (Manakin basiert)

Zur Kurzanzeige

dc.contributor.advisor Schiefer, Ulrich (Prof. Dr. med) de_DE
dc.contributor.author Bruckmann, Anna de_DE
dc.date.accessioned 2007-12-12 de_DE
dc.date.accessioned 2014-03-18T09:39:17Z
dc.date.available 2007-12-12 de_DE
dc.date.available 2014-03-18T09:39:17Z
dc.date.issued 2007 de_DE
dc.identifier.other 275241211 de_DE
dc.identifier.uri http://nbn-resolving.de/urn:nbn:de:bsz:21-opus-31261 de_DE
dc.identifier.uri http://hdl.handle.net/10900/45148
dc.description.abstract Ziel dieser Studie war, die Gesichtsfeld-Untersuchungsergebnisse von zwei verschiedenen Perimetern an Patienten mit fortgeschrittenen Gesichtsfeldausfällen zu vergleichen. Dazu wurde ein Standardgerät, das Octopus 101 (Haag-Streit) und das zu testende, in der Augen-klinik neu entwickelte tragbare Gerät, das Tübingen Mobilcampimeter (TMC) verwendet. Methoden: 31 Patienten im Alter zwischen 18 und 75 Jahren mit fortgeschrittenen Gesichtsfeldausfällen (fortgeschrittene Bogenskotome, konzentrische Einengung, Gesichtsfelddefekte, die die Mittellinie respektieren) und 6 gesunde Probanden wurden in die Studie eingeschlossen. Die erste Untersuchung fand immer am Referenzinstrument Octopus 101 statt: 30° NO Raster mit 192 Stimuli, 10 cd/m² Hintergrundleuchtdichte, Stimulusgröße Goldmann III = 26' und 320 cd/m² maximaler Stimulusleuchtdichte. Die zweite Untersuchung am TMC erfolgte innerhalb der nächsten drei Stunden und verblindet hinsichtlich der Referenzdiagnose mit 84 präsentierten Stimuli als Teil des Octopus 101 Rasters, Hintergrundleuchtdichte zwischen 8 und 20 cd/m², Stimulusgröße 34’ und Stimulusleuchtdichte 320 bis 370 cd/m². Die Genauigkeit (Anteil der an beiden Geräten konkordant wahrgenommenen und konkordant nicht wahrgenommenen Gesichtsfeldorte), die Sensitivität und die Spezifität wurden pro Person ausgezählt und ihre Mediane über die durchschnittlichen Logits berechnet. Ebenso in ein 95%-Konfidenzintervall (KI) geschätzt wurde das geometrische Mittel des Verhältnisses der Untersuchungszeiten. Ergebnisse: Da alle Gesichtsfelder mit dem TMC den gleichen Defektklassen zugeordnet wurden, betragen die Untergrenzen von KI für die Genauigkeit 91 %, die Sensitivität 89 % und die Spezifität 54 % der Personen. Bezüglich der gesamten Stichprobe war der Anteil der am TMC aber nicht am Octopus 101 wahrgenommenen Punkte von allen diskordant wahrgenommenen Punkten 35 % (KI: 30 % bis 40 %). Im Hinblick auf die einzelnen Gesichtsfelddefekte war die mediane punktweise Genauigkeit am größten bei den gesunden Probanden (97,9 %; KI: 97 bis 98.5 %), und am geringsten bei Patienten mit Bogenskotomen (80,6 %; KI: 77,3 % bis 83,5 %). Die mediane punktweise Sensitivität war am höchsten bei Patienten mit konzentrischer Einengung (94.5 %; KI: 82,9 bis 98,4 %), am niedrigsten bei gesunden Probanden (59,1 %; KI: 26,3 bis 85,3 %). Die mediane punktweise Spezifität lag am höchsten bei gesunden Probanden (98,1 %; KI: 96,6 bis 98,9 %), am niedrigsten bei Patienten mit konzentrischer Einengung (77,4 %; KI: 62,1 bis 87,7 %). Die durchschnittliche Untersuchungszeit lag bei 4,6 min am TMC und bei 9,8 min am Octopus 101. Zusammenfassung: Die Ergebnisse zeigen, dass das TMC ein geeignetes Gerät ist, um Gesichtsfeldausfälle sicher aufzudecken und zu dokumentieren. de_DE
dc.description.abstract Purpose: To compare the results of the visual field testing of advanced visual field defects (VFDs) measured with a conventional perimeter (Octopus 101; O-101) and a new portable testing device, the Tuebingen Mobile Campimeter (TMC). Methods: Thirty seven subjects (18 to 75 years old), 13 with advanced arcuate scotomas, 12 with VFD respecting the vertical meridians, 6 with concentric constriction and 6 healthy individuals were included. Subjects were first tested with the Octopus 101 as reference instrument, 30°-NO grid, 192 stimuli, 10 cd/m² background luminance, stimulus Goldmann size III = 26’ and 320 cd/m² stimulus luminance. This was followed by testing with the TMC the same day with 84 stimuli as a subset of the O-101 grid, stimulus size 34', luminance 320 to 370 cd/m², background luminance 8 to 20 cd/m². Pointwise accuracy (proportion of concordant locations), sensitivity and specificity were estimated into 95% confidence intervals (CI) by averaging individual logits. Both examination durations were compared. Results: The TMC results are highly concordant with the results assessed with the Octopus 101 for all four defect classes. All 37 patterns were correctly recognized by the TMC examiner, so that lower limits of CI were for accuracy 91 %, sensitivity 89 % and specificity 54 %. For the entire sample, the percentage of discordant points, perceived with the TMC but not with the O-101, among all discordant points was 35 % (CI: 30 % to 40 %). Analysed by VFD pattern, the average pointwise accuracy was highest in healthy (97.9 %; CI: 97 % to 98.5 %), and lowest in arcuate scotomas (80.6 %; CI: 77.3 % to 83.5 %). Average pointwise sensitivity was highest in concentric constriction (94.5 %; CI: 82.9 % to 98.4 %), lowest in healthy individuals (59.1 %; CI: 26.3 % to 85.3 %). Average pointwise specificity was highest in healthy (98.1 %; CI: 96.6 % to 98.9 %), lowest in concentric constriction (77.4 %; CI: 62.1 % to 87.7 %). Mean patient examination time was 4.6 min with the TMC and 9.8 min with the Octopus 101. Conclusion: The results indicate that the TMC is a feasible device for documented detection of visual field loss. Key words: Visual field – screening method – perimetry – campimetry – portable – concentric – grid en
dc.language.iso de de_DE
dc.publisher Universität Tübingen de_DE
dc.rights ubt-podok de_DE
dc.rights.uri http://tobias-lib.uni-tuebingen.de/doku/lic_mit_pod.php?la=de de_DE
dc.rights.uri http://tobias-lib.uni-tuebingen.de/doku/lic_mit_pod.php?la=en en
dc.subject.classification Gesichtsfeld , Kampimetrie de_DE
dc.subject.ddc 610 de_DE
dc.subject.other Gesichtsfeld , Perimetrie , Kampimetrie , tragbar , Screening Methode de_DE
dc.subject.other Visual field , screening method , perimetry , campimetry , portable en
dc.title Vergleich der Ergebnisse aus Untersuchungen mit dem Tübinger Mobilkampimeter (TMC) und dem Octopus 101 Perimeter an Patienten mit fortgeschrittenen Gesichtsfeldausfällen de_DE
dc.title Comparison of advanced visual field defects, measured with the Tuebingen Mobile Campimeter (TMC) and the Octopus 101 perimeter en
dc.type PhDThesis de_DE
dcterms.dateAccepted 2006-12-01 de_DE
utue.publikation.fachbereich Sonstige de_DE
utue.publikation.fakultaet 4 Medizinische Fakultät de_DE
dcterms.DCMIType Text de_DE
utue.publikation.typ doctoralThesis de_DE
utue.opus.id 3126 de_DE
thesis.grantor 05/06 Medizinische Fakultät de_DE

Dateien:

Das Dokument erscheint in:

Zur Kurzanzeige